7 Maintaining the product
158
Saveris Pt/PtD radio probe
Radio probe with an external Pt100 probe connection
Feature Values
Probe type Pt100
Measuring range -200 to +600 °C
Accuracy
±0.1 °C (0 to +60 °C)
±0.2 °C (-100 to +200 °C)
±0.5 °C (remaining measuring range)
at 25 °C
Resolution 0.01 °C
Connection 1 Pt100 via mini-DIN socket
Protection class IP 68
Operating
temperature
-20 to +50 °C
Saveris H3/H3D radio probe
Humidity radio probe
Feature Values
Probe type NTC Humidity sensor
Measuring range -20 to +50 °C 0 to 100%RH
2
2
Not for non-dewing atmosphere. For constant use in high humidity (> 80%RH
at 30 °C for > 12 h, > 60%RH at > 30 °C for > 12 h), please contact us via
www.testo.com