EasyManua.ls Logo

TestSafe miniPAT - Specifications; Electrical Resistance Ranges; Test Current and Voltage Specifications

TestSafe miniPAT
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SPECIFICATIONS
ELECTRICAL
Resistance Ranges:
Class I Appliances and Cables
The L-E LED will light red when the resistance between line
to earth on the appliance or cable is lower than 1MΩ ±5%.
The N-E LED will light red when the resistance between
neutral to earth is lower than 1MΩ ±5%.
The earth bond LED will light green when the resistance
between the earth of the appliance or cable to the probe is
lower than 0.3Ω ±5%.
Class II Appliances
The Class II result light will be red if the insulation falls below
2MΩ.
The insulation LED's will ash alternately if the resistance is
between 2MΩ and 1MΩ.
The insulation LED's will be solid red if the insulation is
below 1MΩ.
Test Current:
Line to Earth test current is 5uA maximum.
Neutral to Earth test current 5uA maximum.
Earth to Probe test current is 50mA maximum.
17