750W SERIES DC POWER SUPPLY
PU SERIES
PU6-100 6V/100A PU8-90 8V/90A
PU12.5-60
12.5V/60A PU20-38 20V/38A
PU30-25
3V/25A PU40-19 40V/19A
PU60-12.5
60V/12.5A PU80-9.5 80V/9.5A
PU100-7.5
100V/7.5A PU150-5 150V/5A
PU300-2.5
300V/2.5A PU600-1.3 600V/1.3A
INSTRUCTION MANUAL
© PRINTED IN JAPAN B65-0166-00

Table of Contents

Questions and Answers

  • C
    Christopher SmithSep 23, 2025
    What causes poor load regulation on TEXIO Power Supply with CV LED on?
    • S
      Scott FrazierSep 23, 2025
      If you're experiencing poor load regulation on your TEXIO Power Supply and the front panel CV LED is on, it could be due to improperly connected sensing wires. Ensure that the sense wires are connected according to the user’s manual instructions.
  • D
    Danielle MorenoAug 23, 2025
    How to troubleshoot a TEXIO PU6-100 Power Supply with no output and blank displays?
    • A
      Adam RobinsonAug 25, 2025
      First, check if the AC power cord is defective by checking its continuity and replacing it if necessary. Also, verify that the AC input voltage is within the acceptable range and connect the power supply to an appropriate voltage source.
  • E
    Ethan MccoyAug 25, 2025
    What to do if my TEXIO PU6-100 Power Supply output is present momentarily but shuts off quickly and the display indicates “AC”?
    • R
      Rachel WilliamsAug 25, 2025
      Check the input AC voltage. If the AC source voltage sags when a load is applied, connect the power supply to an appropriate voltage source.
  • W
    Willie JacksonAug 27, 2025
    Why TEXIO Power Supply output voltage will not adjust and front panel CC LED is on?
    • M
      Meredith MooreAug 27, 2025
      The unit may be in constant current mode. Check the current limit setting and the load current.
  • D
    Dennis TylerAug 27, 2025
    What to do if my TEXIO Power Supply output is present momentarily but shuts off quickly and the display indicates “OUP”?
    • M
      melaniebrownAug 27, 2025
      The power supply might be configured to Remote sense. Check if the positive or negative load wire is loose.
  • B
    Brian JohnsonSep 18, 2025
    Why does my TEXIO Power Supply have no output and blank displays?
    • G
      Gabriel SanchezSep 18, 2025
      If your TEXIO Power Supply shows no output and all displays and indicators are blank, there could be a couple of reasons: * The AC power cord might be defective: Check its continuity and replace if necessary. * The AC input voltage might be out of range: Verify the input AC voltage and connect to an appropriate voltage source.
  • S
    Sara AllenSep 15, 2025
    Why won't the output current adjust on my TEXIO Power Supply when the CV LED is on?
    • C
      Christopher FryeSep 15, 2025
      If the output current on your TEXIO Power Supply won't adjust and the front panel CV LED is on, it indicates that the unit is in constant voltage mode. Check the current limit and voltage setting.
  • P
    Patrick HoodSep 16, 2025
    What to do if TEXIO Power Supply output voltage won't adjust and CV LED is on?
    • N
      Nicholas MartinezSep 17, 2025
      If the output voltage on your TEXIO Power Supply won't adjust and the front panel CV LED is on, check whether the output voltage is adjusted above the OVP setting or below the UVL setting. Set OVP or UVL so they will not limit the output.
  • V
    Victoria CollinsSep 6, 2025
    Why does my TEXIO Power Supply shut off quickly and display 'AC'?
    • C
      Christian HoustonSep 7, 2025
      If your TEXIO Power Supply output is present momentarily but shuts off quickly, and the display indicates “AC”, it could be due to the AC source voltage sagging when a load is applied. Check the input AC voltage and ensure it's connected to an appropriate voltage source.
  • S
    Sherri SimmonsSep 3, 2025
    What to do if my TEXIO PU6-100 front panel controls are nonfunctional?
    • M
      Michelle SmithSep 3, 2025
      The power supply might be in Local-Lockout mode. Turn off the AC power and wait until the display turns off. Then, turn on the AC power and press the front panel REM/LOC button.

TEXIO PU6-100 Specifications

General IconGeneral
ModelPU6-100
CategoryPower Supply
Output Power600W
Line Regulation≤0.01% + 2mV
CoolingForced Air Cooling
Output Voltage0 to 6V
Output Current0 to 100A
InterfaceRS-232C
Operating Temperature0 to 40°C
Dimensions215(W) x 130(H) x 350(D) mm

Related product manuals