User Manual
Locator and Locator Plus
Cryogenic Storage Vessel with or without
Ultrasonic Level Monitor
Operation Manual and Parts List LT509X9 (7018141) Rev. 4

Questions and Answers

  • B
    Brent PalmerAug 7, 2025
    Why does the Thermo Scientific Locator 6 Plus Laboratory Equipment level monitor alarm sound intermittently?
    • K
      Kyle GilbertAug 7, 2025
      The intermittent alarm and incorrect liquid level measurement in your Thermo Scientific Laboratory Equipment are likely due to ice or frost buildup on the transducer. To resolve this, remove the cover with the monitor from the vessel and let it sit at room temperature for about two hours, or until no ice is visible on the transducer or tube. Do not use a heat gun, as it can damage the cork. Ensure the lid stands upright on the cork to allow moisture to drain. Wipe moisture from the tube's inside and transducer's face with a clean, soft cloth. Then, replace the cover with the monitor and press the “Reset” button.
  • R
    Ronnie RayAug 20, 2025
    What to do if Thermo Scientific Locator 6 Plus level monitor reads 'Full' after replacing the lid?
    • D
      Deborah StephensAug 20, 2025
      If your Thermo Scientific Laboratory Equipment's level monitor reads 'Full' after you replace the lid, it's probably due to fog in the top part (neck area) of the cryogenic vessel. The density of this fog depends on the room's humidity. The fog should dissipate in about fifteen minutes, after which the level monitor will provide an accurate reading of the liquid nitrogen level.

Summary

Section 1 Safety Information

Please Read

Important initial safety guidance and vessel handling advice.

Warnings

Critical safety warnings regarding cryogenic hazards, personal protection, and handling.

Cautions

Specific precautions to prevent damage and ensure safe operation, including filling and maintenance.

Section 4 Operation

Cautions

Critical precautions for operation, including filling and rack handling.

Filling Instructions

Step-by-step procedure for adding liquid nitrogen to a warm vessel.

Section 6 Liquid Nitrogen Level Monitor Operation

Installation

Steps for installing the liquid nitrogen level monitor.

Level Monitor Remote Alarm

Details on connecting the level monitor to a remote alarm system.

Thermo Scientific Locator 6 Plus Specifications

General IconGeneral
Capacity6 liters
Vial Size2 ml
MaterialAluminum
Neck Diameter5 inches
Operating Temperature-196°C to +30°C
Storage Temperature-196°C to +30°C