Multiskan
EX
is an eight-channet vertical light path filter photometer designed to pelrom both standard
photometric. and agglutination measurements.
There
is
a versatile built-in software available in Multiskan
EX.
It
comprises several software
(program modules). Each package is a combination of measurement and calculation modes.
memory capacity is achieved by programmable program card
The
wavelensh
(400
-
750
nm or optional
340
-
450
nm) is selected using mmirnum eight high quality
interference filters held in a
filter
wheel.
processed via a serial
RS-232C
interface or via
a
parallel Centronics
The optional huilt-in thermal printer provides also a record ofthe results.
Multiskan
EX
utilizes the original Lahystems' concept of vertical photometry in which the light beam passes
through the whole sample.
In vertical photometry, the absorption of light
is
proportional to the amount
of
light absorbing substance
in
the
well.
Absorption is expressed by the following equation:
where A
=
absorbance, a
=
molar absorptivity of
the
substance,
m
=
mass of absorbing substam and
S
=
cms-
ndimlar to the light path.
The
advantages of
lhe
vertical light path measurement are:
(I)
inaccurate pi ning of nonabsorbing lrquids does not affect the measured absorbance values
(2)
evaporation of nonabsorbing liquids during the reaction
dm5
not atfect the measured absorbance values
and
(3)
a cenaiain degree of inhomogeneity in the solution,
for
example
as
a result of layering in turbidity
measurements,
does
not
affect
the results.