41
OPTIONAL ACCESSORIES
OPTIONAL ACCESSORIES
PACHOMETER ATTACHMENT
Used for measuring the corneal thickness.
* Two LED alignment targets are projected onto the cornea to assure repeatable alignment of
the slit beam. With the Mishima-Hedby method, the slit beam can be accurately positioned
perpendicular to the corneal surface.
* A separate LED illumination is placed near the scale to enhance reading, even in a darkened
environment.
* A magnifier is placed over the scale for easier reading.
10× MEASURING EYEPIECE
When this accessory is used in place of the normal eyepiece, linear and angle measurements
become possible. It is used for the fitting of toric contact lenses as well.
Scale specifications
Linear Scale: 16mm
0.5mm minimum graduations
Angle Scale: 360°
5° minimum graduations
Measuring parameters
Linear scale : To be used at 10× magnification only.
Diopter compensation range: -5D to +3D
Angle Scale : No limitations
Pachometer
Eyepiece