62
BEFORE REQUESTING SERVICE
* A part of pixels might keep ON or OFF due to the characteristics of the color LCD monitor, but it
is not a trouble. Please use the instrument as it is. This phenomenon does not affect the image
data.
Split lines cannot be
seen.
Split line is set to OFF. Turn split line ON (Split switch).
13
Diopter compensation lens selector is
not set to (0).
Return diopter compensation lens selector
to (0).
42
Patient's pupil is not large enough. Darken room and thoroughly dilate
patient's eye.
44
Xenon lamp does not
turn on.
Power save function is on (flickering
power lamp).
Press Photography switch and cancel
power save function.
36
Xenon lamp has served its life. Change xenon lamp.
72
Anomaly in external recording device. Check power supply, settings, paper, etc.
Cannot get patient's
pupil center.
Is the patient's chin and forehead posi-
tioned correctly on the forehead rest and
chinrest?
Is the patient facing sideways?
Have patient correct their position.
39
Patient's face height is incorrect. Adjust the face height with the chinrest
switch.
39
Nothing is recorded in
external recording
device.
Anomaly in external recording device. Check power supply, settings, paper, etc.
Cable connections are incorrect. Check and correct cable connections.
24
Problem Condition Check Page