27
10. Adjusting Zero Position and Sensitivity with Zero
and Sensitivity Adjustment Knobs
The zero position and sensitivity of the air circuit need not be adjusted, except after a
measuring head change or after the output of the mastering result NG due to wear.
10.1 Adjusting
In the measure mode, press for 2 sec or more.
The MAS LED below the bargraph display comes on, and “MAX M.” or “MIN M.”
appears on the alphanumeric display. Press to change to the adjust
detector mode. The alphanumeric display shows “ADJ.” for 2 sec and then the
measured value.
Fig. 10-1. Changing to adjust detector mode.
2
1
3
SET
MAS
ENT
RST
SET MAS
2
1
3
SET
MAS
ENT
RST
SET MAS
MIN M.
12.3
Press 2 sec or more
Press
2
1
3
SET
MAS
ENT
RST
SET MAS
ADJ.