7. Mastering
7.1 Automatic mastering
In the measure mode, press for 2 sec or more.
The MAS LED below the bargraph display comes on orange, “MIN M.” appears on the
alphanumeric display, and the instrument changes to the master mode.
Fig. 7-1. Changing to master mode.
Set the minimum master in the measuring nozzle, and press . The instrument is
calibrated with the minimum master, and “MAX M.” is shown on the alphanumeric
display.
Fig. 7-2. Calibrating with minimum master.
2
1
3
SET
MAS
ENT
RST
SET MAS
2
1
3
SET
MAS
ENT
RST
SET MAS
12.3
MIN M.
Press 2 sec or
more
ENT
RST
SET MAS
ENT
RST
SET MAS
Set minimum master in
measuring nozzle, and
press this
MAX M.
Completion of
minimum-mastering
Measure mode Master mode