29
Set the minimum master in the measuring head, and turn the zero adjustment
knob to set the indication at the minimum master value.
Fig. 10-3. Adjusting zero position.
Set the maximum master in the measuring head. If the indication is about the
maximum master value, the adjustment is finished.
Fig. 10-4. Finishing adjustment with zero and sensitivity adjustment knobs.
When the indication is not in the automatic mastering range (or the bargraph
display is illuminated red), refer to “10.2 Indication below mastering range” on
page 31 or “10.3 Indication above mastering range” on page 32.
ZERO
MAG
GAGE
ENT
RST
SET MAS
CAG
2000
Set minimum master
in measuring head
and turn knob
1
3
SET
MAS
2
Set indication at
minimum master
value
Adjustment
finished