EasyManua.ls Logo

Townsend Labs Sphere L22 - OAC Performance and Patterns

Townsend Labs Sphere L22
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Townsend Labs Sphere L22 Precision Microphone System Guide Off-Axis Correction
34
Figure 5-2 compares the frequency response from directly behind a U87ai mic (180° off-axis, 1 m
from the source, with
1
/
6
-octave smoothing) with a Sphere mic (same position) with Off-Axis
Correction enabled. The green line shows the frequency response of a Neumann U87ai; the blue line
shows the response of the Sphere mic. The Sphere polar response is much flatter: The U87 response
varies by as much 16 dB, whereas the Sphere mic varies by about 6 dB with smoother overall
fluctuations. Not surprisingly, the U87ai’s audio from 180° off-axis sounds very colored, while the
Sphere produces a smoother sound comprised mostly of natural room ambience.
Figure 5-2 U87ai Cardioid 180° off-axis vs. Sphere Off-Axis Correction
Another potential issue is that the on-axis frequency response of virtually all multi-pattern mics also
changes dramatically when adjusting the polar pattern. This can be a problem when you have the
sound dialed in, but then later decide to change the pattern to reduce bleed or feedback. With Off-
Axis Correction, it is now possible to adjust the polar pattern without changing on-axis frequency
response.
Figure 5-3 shows the on-axis frequency response of a U87ai at its three pattern settings. At low and
high frequencies, omni and figure-8 differ in response by up to 8 dB.

Table of Contents