EasyManuals Logo

TPS pH Cube Handbook

Default Icon
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
Page 14
7.2.3 Temperature Compensation
The slope of a pH sensor (section 7.2.2) is affected by temperature. This effect is
compensated for either by using an Automatic Temperature Compensation
(ATC) sensor or by entering the sample temperature manually. Figure 7-3 shows
the slope of a pH sensor at various temperatures.
-600
-400
-200
0
200
400
600
0714
Electrode
Potential (mV) at
0 oC (54mV/pH)
Electrode
Potential (mV) at
50 oC (64mV/pH)
Electrode
Potential (mV) at
100 oC (74mV/pH)
pH Electrode Response, as a Function of Temperature
Electrode Response (mV)
pH
Figure 7-3
7.3 Checking the reference junction of a pH sensor.
If pH readings are inaccurate or unstable, the reference junction of the sensor
may be blocked. The following test can be performed to determine if the
reference junction of a pH sensor is making adequate contact with the sample
solution.
1. Calibrate the pH Cube, as per section 3.
2. Dilute 1 part of pH6.88 buffer with 9 parts of distilled water.
3. Measure the pH of the diluted buffer. The result should be 7.06 +/-0.05 pH.
4. If the value obtained is outside of these limits, then clean the reference
junction as per the instructions supplied with the pH sensor.
5. Re-calibrate the pH Cube and repeat the test.
6. If the value obtained is still outside 7.06 +/-0.05 pH, then the sensor should be
replaced.

Other manuals for TPS pH Cube

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TPS pH Cube and is the answer not in the manual?

TPS pH Cube Specifications

General IconGeneral
BrandTPS
ModelpH Cube
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals