Congratulations !
Your new miniCHEM-pH monitor is the latest in simple, reliable process
monitoring instrumentation. With correct operation and maintenance, your
miniCHEM-pH will give you many years of reliable service.
The miniCHEM-pH is a breeze to operate. This manual has been designed to
help you get started, and also contains some handy application tips. If at any
stage you require assistance, please contact either your local TPS representative
or the TPS factory in Brisbane.
The manual is divided into the following sections:
1. Table of Contents
Each major section of the handbook is clearly listed. Sub-sections have also
been included to enable you to find the information you need at a glance.
2. Introduction
The introduction has a diagram and explanation of the display and controls of
the miniCHEM-pH. It also contains a full listing of all of the items that you
should have received with the unit. Please take the time to read this section, as
it explains some of items that are mentioned in subsequent sections.
3. Main Section
The main section of the handbook provides complete details of the
miniCHEM-pH, including operating modes, calibration, troubleshooting,
specifications, and warranty terms.
4. Appendices
Appendices containing background information and application notes are
provided at the back of this manual.
miniCHEM-pH
Process Monitor
Version : 2.11
Date : 10/09/2002
Author : MS

Questions and Answers

  • J
    Jason AndersonAug 15, 2025
    How to fix no display on TPS Measuring Instruments?
    • J
      john59Aug 15, 2025
      If your TPS Measuring Instrument's display is not working, first, ensure the mains power is switched ON. Next, check the power connections to make sure they are correctly connected. If the problem persists, the instrument may be faulty and should be returned to TPS for repair.
  • J
    josephwelchAug 18, 2025
    Why are the readings unstable on my TPS miniCHEM-pH?
    • K
      kbarnettAug 18, 2025
      Unstable readings on your TPS Measuring Instruments can occur due to several reasons: 1. The solution guard may not be fitted properly; connect the green wire from the ATC (or solution guard rod if no ATC sensor is fitted), to terminal 3. 2. The reference junction or glass bulb may be dirty; clean them as per instructions supplied with the electrode. 3. There might be a bubble in the glass bulb; flick the electrode to remove it. 4. The reference junction may not be immersed; ensure that the bulb AND the reference junction are fully immersed. 5. KCl crystals may be present around reference junction inside the electrolyte chamber; rinse the electrolyte chamber with warm distilled water until dissolved and replace the electrolyte.
  • C
    Cheyenne FullerAug 21, 2025
    What to do if TPS miniCHEM-pH Measuring Instruments have an incorrect analogue output signal?
    • Y
      ymeadowsAug 21, 2025
      If your TPS Measuring Instruments are showing an incorrect analogue output signal, check the following: 1. Verify that the Current/Voltage Output jumpers are correctly set for 4-20mA or 0-1V, and adjust if necessary. 2. Ensure the 4-20mA loop resistance does not exceed 500 Ohms. 3. Make sure the load on the 0-1V DC output is a minimum of 1000 Ohms. 4. Calibrate the 4-20mA output. If the issue persists, the instrument may be faulty and should be returned to TPS for repair.
  • K
    kentstanleyAug 25, 2025
    Why does the span calibration fail on my TPS miniCHEM-pH?
    • C
      Catherine CochranAug 26, 2025
      A span calibration failure on TPS Measuring Instruments with insufficient range can occur due to: 1. The ATC probe not being connected; connect it to terminals 1 & 2. 2. A dirty glass bulb; clean it as per instructions supplied with the electrode. 3. An aged electrode; attempt rejuvenation as per instructions supplied with the electrode, or replace it if necessary. 4. Inaccurate buffers; replace them.
  • H
    Haley JohnstonAug 28, 2025
    Why doesn't the alarm LED work on my TPS Measuring Instruments?
    • D
      Diana MillerAug 28, 2025
      If the alarm LED or optional relay output on your TPS Measuring Instruments does not operate when the limit is exceeded, it could be due to the following: 1. The limit may not be set correctly for “too high” or “too low” alarm/control; set the Lo/Hi Output jumpers correctly. 2. The optional relay output may not be fitted; return the unit to the factory to have the relay output fitted (part no 130342). 3. The hysteresis may be too large; check and adjust the HYSTERESIS trimmer, setting it to the minimum level if unsure. 4. A blown fuse; check and replace the fuse if necessary. If the problem continues, the instrument may be faulty and should be returned to TPS for repair.
  • J
    Joshua CarterAug 31, 2025
    What to do if TPS miniCHEM-pH show inaccurate readings after calibration?
    • T
      Tabitha EwingSep 1, 2025
      If your TPS Measuring Instruments provide inaccurate readings even after successful calibration, the reference junction may be blocked. Clean the reference junction as per the instructions supplied with the electrode.
  • B
    Brian KaneSep 4, 2025
    What causes asymmetry calibration failure in TPS miniCHEM-pH?
    • M
      Matthew RiveraSep 4, 2025
      Asymmetry (Zero) calibration failure with insufficient range on your TPS Measuring Instruments can be caused by: 1. A blocked reference junction; clean it as per instructions supplied with the electrode. 2. Contaminated reference electrolyte; flush with distilled water and replace the electrolyte.
  • R
    Rebecca GardnerSep 7, 2025
    Why do TPS miniCHEM-pH Measuring Instruments display around 7 pH in all solutions?
    • K
      Kathryn LopezSep 7, 2025
      If your TPS Measuring Instruments consistently display around 7 pH in all solutions, there may be an electrical short inside the electrode or cable. Test the system with a new electrode and replace the old one if necessary.
  • S
    sguzmanSep 10, 2025
    What to do if TPS Measuring Instruments display 4-5 pH for all solutions?
    • J
      jacqueline52Sep 12, 2025
      If your TPS Measuring Instruments display 4-5 pH for all solutions, the glass bulb or internal stem may be cracked. Replace the electrode.

Summary

Introduction

miniCHEM-pH Illustration

Detailed diagram and explanation of the miniCHEM-pH front panel controls and display.

Unpacking Information

Lists the standard accessories included with the miniCHEM-pH and optional items.

Specifications

Comprehensive list of technical specifications for the miniCHEM-pH monitor.

Main Section

Operating Modes

Explains the function of each operating mode selectable via the function switch.

Installation and Set-up

Guides through physical installation, wiring, and initial configuration of the miniCHEM-pH.

Calibration

Details the step-by-step procedures for zero and span calibration of the pH sensor.

Process Monitoring

Describes how to utilize the miniCHEM-pH for continuous monitoring of process parameters.

Automatic Dosing

Explains how to use the unit for automatic control and alarming based on pH limits.

Manual Dosing

Instructions for manually activating dosing functions using the unit's controls.

Fuse Replacement

Procedure for safely replacing the protective fuse within the miniCHEM-pH unit.

Troubleshooting

Provides common issues and solutions for instrument and pH related problems.

Warranty

Outlines the warranty terms, conditions, and procedure for service requests.

Appendices

Checking the reference junction of a pH electrode

A test procedure to verify the functionality of the pH electrode's reference junction.

Determining if an instrument or electrode is faulty

Diagnostic steps to identify whether the instrument or the pH electrode requires servicing.

Terminal connections when 12 V DC option is fitted

Specific terminal connection details for units equipped with the 12V DC power option.

Relay Output Wiring Examples for mains powered units

Wiring diagrams illustrating the connection of the optional relay output for mains powered units.

Relay Output Wiring Examples for 12 V DC powered units

Wiring diagrams illustrating the connection of the optional relay output for 12V DC powered units.

Drilling Template

A full-size template for accurately drilling mounting holes for the miniCHEM enclosure.

TPS miniCHEM-pH Specifications

General IconGeneral
pH Resolution0.01 pH
Temperature Resolution0.1 °C
Temperature Accuracy±0.5 °C
CalibrationAutomatic, up to 3 points
Battery LifeUp to 500 hours
DisplayLCD
Temperature CompensationAutomatic
pH Range0 to 14 pH
Measurement Range0 to 14 pH