Congratulations !
You have purchased the latest in handheld Conductivity-Salinity-pH-
Temperature instrumentation. We trust that your new MC-81 will give you
many years of reliable service.
The MC-81 is a breeze to operate. This manual has been designed to help you
get started, and also contains some handy application tips. If at any stage you
require assistance, please contact either your local TPS representative or the
TPS factory in Brisbane.
The manual is divided into the following sections:
1. Table of Contents
Each major section of the handbook is clearly listed. Sub-sections have also
been included to enable you to find the information you need at a glance.
2. Introduction
The introduction has a diagram and explanation of the display and controls
of the MC-81. It also contains a full listing of all of the items that you
should have received with your MC-81. Please take the time to read this
section, as it explains some of items that are mentioned in subsequent
sections.
3. Main Section
The main section of the handbook provides complete details of the MC-81,
including operating modes, calibration, troubleshooting, specifications, and
warranty terms.
4. Appendices
Appendices containing background information and application notes are
provided at the back of this manual.
TPS Pty Ltd
4 Jamberoo Street
Springwood, Brisbane,
Australia, 4127
Phone : (07) 32 900 400
International : 61 7 32 900 400
Fax : (07) 3808 4871
International : 61 7 3808 4871
Email : tps@tpsite.com.au
Web Site : www.tpssite.com.au
Model MC-81
Conductivity-Salinity-
pH-Temp. Meter
Handbook Version : 1.1
Date : 16-Sep-99
Author : MS

Questions and Answers

  • J
    Jeffrey RothSep 23, 2025
    What does 'Factory Cal. Fail See Handbook' mean on my TPS Measuring Instruments?
    • C
      christine23Sep 23, 2025
      If your TPS Measuring Instruments displays 'Factory Cal. Fail See Handbook', it means that the EEPROM chip which contains the factory calibration information has failed. The unit must be returned to TPS for service. Conductivity and Salinity readings will be accurate after a full two point calibration. pH readings will be accurate after a 2-point calibration (use manual temperature compensation). Temperature readings may be up to ±5oC incorrect.
  • V
    veronica85Aug 23, 2025
    What to do if my TPS Measuring Instruments meter will not turn on?
    • E
      Eric LuceroAug 23, 2025
      If your TPS Measuring Instruments meter will not turn on, it's likely because the battery is exhausted. Replacing the battery should resolve the issue.
  • N
    nelsonmichaelAug 25, 2025
    What does the flashing symbol on my TPS MC-81 mean?
    • G
      Gina RodriguezAug 26, 2025
      If your TPS Measuring Instruments shows a flashing symbol, it indicates that the battery voltage is below 7.50 volts. You should replace the battery soon. The unit will automatically switch off when the battery falls below 6.00 volts.
  • J
    james18Aug 27, 2025
    Why are my TPS MC-81 giving inaccurate readings even after calibration?
    • L
      Laura WoodsAug 27, 2025
      If you get inaccurate readings with your TPS Measuring Instruments, even after successful calibration, it could be due to a build-up of dirt or oily material on the electrode wires, or the platinum-black coating may have worn off. You should clean the electrode. If the platinum-black coating is worn off, the electrode requires replatinisation.
  • C
    Crystal PetersAug 28, 2025
    Why are the readings on my TPS Measuring Instruments low or near zero?
    • S
      Shaun Wilcox MDAug 28, 2025
      If your TPS Measuring Instruments readings are low or near zero, it could be due to a build-up of dirt or oily material on the electrode wires, the electrode not being immersed deeply enough, or a faulty electrode. Clean the electrode. Ensure the electrode is immersed at least to the vent hole in the white plastic cover. If the problem persists, the electrode may need to be returned to the factory for repair or replacement.
  • K
    kevinmclaughlinSep 18, 2025
    What does 'Memory Failed Calib. Lost Memory Reset' mean on my TPS MC-81 Measuring Instruments?
    • J
      Jeremy TaylorSep 18, 2025
      If you see 'Memory Failed Calib. Lost Memory Reset ! You MUST Re-Cal.' on your TPS Measuring Instruments, it means that the user calibration settings have been lost or corrupted. Re-calibrate the instrument. Both a Zero and a Standard calibration will be required for Conductivity and Salinity. A full 2 point calibration will be required for pH and a 1 point calibration for temperature.
  • V
    Victoria RichardsSep 15, 2025
    Why are the readings on my TPS MC-81 Measuring Instruments unstable?
    • I
      Ian SerranoSep 15, 2025
      If you are experiencing unstable readings on your TPS Measuring Instruments, it could be due to several reasons: a blocked reference junction, a dirty glass bulb, a bubble in the glass bulb, a faulty connection to the meter, the reference junction not being immersed, or KCl crystals around the reference junction inside the electrolyte chamber. Clean the reference junction. Clean the glass bulb. Flick the electrode to remove any bubble. Check the connectors and replace if necessary. Ensure that the bulb AND the reference junction are fully immersed. Rinse the electrolyte chamber with warm distilled water until the crystals are dissolved, then replace the electrolyte.
  • L
    lovestevenSep 16, 2025
    Why is the temperature inaccurate on my TPS MC-81 Measuring Instruments and cannot be calibrated?
    • J
      Jennifer RoblesSep 17, 2025
      If the temperature is inaccurate on your TPS Measuring Instruments and cannot be calibrated, it could be due to a faulty connector, a faulty temperature sensor in the conductivity sensor, or a faulty instrument. Check the connector and replace if necessary. Fit a new conductivity sensor. Note that the conductivity readout may still be accurate. If the problem persists, return the instrument to the factory for repair.
  • D
    diazjanetSep 6, 2025
    What causes a 2 point calibration failure on my TPS Measuring Instruments?
    • C
      Christopher ArnoldSep 7, 2025
      If the 2 Point calibration fails on your TPS Measuring Instruments (Slope is less than 85.0%), it could be due to several reasons: using an incorrect primary buffer, the glass bulb not being clean, the electrode being aged, the connector being damp, or inaccurate buffers. Ensure that you are using the buffers which the MC-81 has been set to automatically recognise. Clean the glass bulb. Attempt rejuvenation. If not successful, replace the electrode. Dry the connector in a warm place. Replace the buffers.
  • D
    Diane ThomasSep 3, 2025
    What causes a 1 point calibration failure on my TPS Measuring Instruments?
    • A
      Ashley OlsonSep 4, 2025
      If the 1 Point calibration fails on your TPS Measuring Instruments (Asymmetry is greater than +/-1.00 pH), it may be because the reference junction is blocked or the reference electrolyte is contaminated. Clean the reference junction and flush with distilled water and replace the electrolyte.

Summary

Introduction

MC-81 Display and Controls

Explains the display and controls of the MC-81 instrument.

Unpacking Information

Lists accessories included with the MC-81 instrument.

Specifications

Details the technical specifications of the MC-81 instrument.

Display Formats and Readout Modes

MC-81 Display Formats

Describes the two available display formats: Dual and Large.

Readout Modes for Dual Display

Details the various readout modes available when using the Dual Display format.

Readout Modes for Large Display

Details the readout modes specifically for the Large Display format.

Conductivity Calibration

Calibration Procedure

Step-by-step guide for calibrating conductivity measurements.

Conductivity Calibration Notes

Provides important notes and recommendations for conductivity calibration.

Conductivity Calibration Messages

Explains the messages displayed during conductivity calibration.

Salinity Calibration

Calibration Procedure

Step-by-step guide for calibrating salinity measurements.

Salinity Calibration Notes

Provides important notes and recommendations for salinity calibration.

Salinity Calibration Messages

Explains the messages displayed during salinity calibration.

pH Calibration

Calibration Procedure

Step-by-step guide for calibrating pH measurements.

pH Calibration Notes

Provides important notes and recommendations for pH calibration.

pH Calibration Messages

Explains the messages displayed during pH calibration.

Temperature Calibration

Temperature Calibration

Guide on how to calibrate the temperature sensor.

Calibration Notes

Notes regarding temperature calibration procedures.

Calibration Messages

Explains messages related to temperature calibration success or failure.

Manual Temperature Setting

Instructions for manually setting the temperature.

Battery Saver Function

Selecting Buffers for Auto Buffer Recognition

Initialising the MC-81

Troubleshooting

General Error Messages

Lists common error messages, their causes, and remedies.

Conductivity and Salinity Troubleshooting

Addresses common issues encountered during conductivity and salinity measurements.

pH and mV Troubleshooting

Addresses common issues encountered during pH and mV measurements.

Temperature Troubleshooting

Addresses common issues with temperature readings and calibration.

Warranty

Appendices

Soil and Water Salinity Information

Information on soil salinity, measurement methods, and water classification.

Electrode Care, Maintenance & Fundamentals

Covers care, maintenance, replatinising of conductivity electrodes, and pH electrode basics.

pH Electrode Troubleshooting & Instrument Info

Focuses on checking pH electrode junctions, diagnosing faults, and finding software version.

TPS MC-81 Specifications