6.3 Calibration Messages
1. If a temperature calibration has been successfully performed, the MC-81 will
display the following message and then the offset value of the probe.
eg: Temp then: Offset
Cal. OK 1.0
o
C
2. If a temperature calibration has failed, the MC-81 will display the following
message, and then the failed offset value of the probe.
eg: Temp then: Offset
Cal.Fail 10.5
o
C
6.4 Manual Temperature Setting
1. Press the key if the meter is not already switched on.
2. Press the key until the meter is in Temperature mode.
3. Manual temperature setting is only available if the Conductivity sensor is not
connected.
4. Press and hold the key for 2 seconds.
5. The display should now look like this:
Man.Temp
↑↑ 25.0↓↓
6. Press the and keys until the bottom line shows the temperature which
you wish to set. This value should be the same as the temperature of the
solution you are measuring.
Press the key to set the temperature.