EasyManua.ls Logo

TPS MC-81 - pH Electrode Troubleshooting & Instrument Info

TPS MC-81
37 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 35
12.6.3 Temperature Compensation
The slope of a pH electrode (section 12.6.2) is affected by temperature. This
effect is compensated for either by using an Automatic Temperature
Compensation (ATC) probe or by entering the sample temperature manually.
Figure 12-3 shows the slope of a pH electrode at various temperatures.
-600
-400
-200
0
200
400
600
0 7 14
Electrode
Potential (mV) at
0 oC (54mV/pH)
Electrode
Potential (mV) at
50 oC (64mV/pH)
Electrode
Potential (mV) at
100 oC (74mV/pH)
pH Electrode Response, as a Function of Temperature
Electrode Response (mV)
pH
Figure 12-3
12.7 Checking the reference junction of a pH electrode.
If pH readings are inaccurate or unstable, the reference junction of the electrode
may be blocked. The following test can be performed to determine if the
reference junction of a pH electrode is making adequate contact with the sample
solution.
1. Calibrate the MC-81, as per section 5.1.
2. Dilute 1 part of pH6.88 buffer with 9 parts of distilled water.
3. Measure the pH of the diluted buffer. The result should be 7.06 +/-0.05 pH.
4. If the value obtained is outside of these limits, then clean the reference
junction, as per the instructions supplied with the pH electrode.
5. Re-calibrate the MC-81 and repeat the test.
6. If the value obtained is still outside 7.06 +/-0.05 pH, then the electrode should
be replaced.

Related product manuals