Page 16
7. Battery Saver Function
The MC-81 is equipped with a battery saver function. If no button has been
pressed for five minutes, the unit beeps and flashes the display for 20 seconds,
and then shuts off. This function can be disabled for continuous use.
To enable or disable the battery saver function:
1. Press the key if the meter is not already switched on.
2. With the meter already switched on, press and HOLD the key for 3
seconds.
3. The display should now look like this:
Either: :ON or: :OFF
9.00V 9.00V
4. In this mode, use the or keys to toggle the battery saver function on or
off.
NOTE: The display also shows the battery volts. This gives the operator an idea
of how much battery life is remaining. The symbol flashes when the
battery volts drops below 7.50 volts. At 6.00 volts the meter turns itself
off.
5. When you have set the battery saver function to the desired position, press the
key to return to normal measurement mode. The setting is kept in memory
when the meter is switched off, even if the battery is removed.