3200B SERIES OPERATION MANUAL
TRANSMILLE LTD. Page 73
Calibration can be carried out automatically via the USB interface if required.
Adjustments to the calibration can only be made using the interface see remote
commands section of this manual for details. A Calibration Control Panel (CCP)
program is available from Transmille which allows full control and adjustment of the
calibrator.
The recalibration of 3200B calibrator should be performed annually in a standards
laboratory with the correct equipment. Adjustment should not be attempted without
the required standards.
Before calibration it is important to have met the conditions listed below
1: The correct environmental condition.
2: The calibrator must have fully warmed up and been allowed to stabilise for the
correct period of time.
3: Operated from the correct line voltage.
4: To use the correct calibration equipment.
5: To have available the Correct Test leads required
6: To understand the required test specifications.
7: To operate the calibrator at all times within its load and voltage capabilities
COMPREHENSIVE CALIBRATION INSTRUCTIONS ARE AVAILABLE IN THE
3200B SERVICE KIT (OPTION).