EMC GUIDANCE
EMC GUIDANCE
Guidance and manufacturer’s declaration – electromagnetic Immunity
Immunity Test
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±8 kV contact
Table 2
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
IEC 60601-1-2
Test level
±2 kV, ±4kV, ±8 kV,
±15 kV air
Not application
Not application
Not application Not application
Not application
Not application
30 A/m
50Hz/60Hz
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
30 A/m
50Hz/60Hz
Power frequency
magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Radiated RF
IEC61000-4-3
Not application
Not application