Mode d’emploi / User manual TOM 011
15 / 18
Mesureur Puissance Optique
Optical Power Meter
TOM 011 – Référence / Part Number: 307967
Setting Process
H. After opening the battery compartment and taking out the batteries, you
will find a little red switch with a white button on it and push the white
button to “on” position. Then please put the batteries and compartment
back and open the unit to re-calibrate it.
I. Press the “REF” key once to increase the current power by 0.05dB
which will be displayed on the LCD screen.
J. Press the “dBm/w” key once to decrease the current power by 0.05dB
which will be displayed on the LCD screen.
K. Press¬ākey to switch the working wavelength.
L. Press ĀON/OFFākey shortly to store the self-calibrating value.
M. Press ĀON/OFFākey for 2 seconds to turn off the unit.
N. Take out the battery and push that white button to “OFF” position to exit
the self-calibrating mode.
3. Maintenance
z Clean the optical connector and adapter regularly.
z Use the regulated optical connector for the test.
z Shut off the power and cover the dust-proof cap after your test.
z Select the power supply within the required voltage range when using
AC adapter.
z Please take out the battery when it is not in operation for a period of time.