3.6 Manual selection of Exposure time
The user can manually modify the pre-set exposure times for any type of receptor, pressing one of
the buttons “+” or “-“ from the main keyboard or on the optional keyboard. Times can move of 47
steps from 0.010 to 2s as per the following table:
4 CARE AND MAINTENANCE
4.1 Cleaning
Always disconnect the line voltage supply before cleaning the unit. Use a mild soap to remove fin-
ger or other dirty marks paying attention not to have liquids enter into the equipment.
Clean painted surfaces , accessories and connection cables only by hand with
a damp cloth and mild detergent, making sure to wipe with a dry cloth ; Do not
use solvents ( alcohol, benzene , trichlorethylene ), corrosive and abrasive.
4.2 Disinfecting
Parts in touch with the patient must be cleaned with a detergent (such as 2% solution of ammonia)
making sure not to use solvents or corrosive disinfectants, which can cause cracks on the plastic
covers.
4.3 Maintenance
Maintenance of RiX- 70 DC to be done regularly by a service technician at least once every 24
months, in addition to regular checks performed by the operator every year.
RiX-70 DC system not contain parts that can be repaired by the user.
The system must be opened and maintained by qualified personnel.