6 EN
vane anemometer BA16
7. Switch the device off.
ð The set value of the cross-sectional area will be used
for the calculation of the volumetric flow after the next
switch-on.
Indication of minimum and maximum values
1. Press the MAX/ MINbutton(7).
ð The RECORD indication(22) appears on the display.
2. Press the MAX/MIN button(7) to switch between the
maximum and minimum value.
ð MAX(18) appears on the display.
ð Actuating the MAX/ MINbutton(7) starts a new
measurement. The value now appearing on the display
is the reference value.
ð MAXfunction: The maximum value of the newly started
measurement interval is displayed. Values smaller than
the one displayed will not be indicated.
3. Press the MAX/MIN button(7) again.
ð MIN(17) appears on the display.
ð Actuating the MAX/ MINbutton(7) starts a new
measurement. The value now appearing on the display
is the reference value.
ð MINfunction: The minimum value of the newly started
measurement interval is displayed. Values exceeding
the one displayed will not be indicated.
4. Press the MAX/ MINbutton(7) for approx. 3seconds to
terminate the function.
ð The values will be deleted from the buffer memory.
Using the HOLDfunction
1. Press the HOLD/lampbutton(5).
ð The current value will be indicated on the display.
2. Press the HOLD/lampbutton(5). again.
ð The display again indicates the current measured value.
Disabling automatic switch-off
As standard, the automatic switch-off function is activated. This
results in the device switching off automatically after
15minutes of non-use. Please proceed as follows to deactivate
the automatic switch-off function:
ü The device is switched off.
1. Additionally press the UNITSbutton(6) when switching the
device on.
ð The indication disAPO appears on the display.
ð Automatic switch-off is deactivated.
Info
Note that the automatic switch-off function is activated
upon the next switch-on. If you do not like to use the
function, deactivate the automatic switch-off function
as described.
Switching the device off
1. Briefly press the Power button(8).
ð The device will be switched off.
ð The saved measured maximum and minimum values
are reset.
Maintenance and repair
Battery change
A battery change is required when the battery status
indication(20) lights up or the device can no longer be switched
on (see chapter Inserting the battery).
Cleaning
Clean the device with a soft, damp and lint-free cloth. Make
sure that no moisture enters the housing. Do not use any
sprays, solvents, alcohol-based cleaning agents or abrasive
cleaners, but only clean water to moisten the cloth.
Repair
Do not modify the device or install any spare parts. For repairs
or device testing, contact the manufacturer.
Errors and faults
The device has been checked for proper functioning several
times during production. If malfunctions occur nonetheless,
check the device according to the following list.
Fault Cause Remedy
The device can no
longer be switched
on.
The battery is dead. Insert a new battery
(see chapter
Operation).
The device indicates
implausible values
of the volumetric
flow measurement.
The cross-sectional
area of the opening
has not been
entered (correctly).
Determine and enter
the cross-sectional
area of the opening
(inm²) (see chapter
Operation).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Trotec BA16 and is the answer not in the manual?

Trotec BA16 Specifications

General IconGeneral
BrandTrotec
ModelBA16
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals