EasyManua.ls Logo

TSA PRM-470B - 5 Maintenance; Maintenance Procedures; Electronic Calibration Procedure

TSA PRM-470B
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. MAINTENANCE page 19
Model PRM-470B
5. MAINTENANCE
5.1. MAINTENANCE PROCEDURES
The PRM-470B has been designed for continuous use up to 50 hours on a set of batteries. Once
initial installation has been completed, little maintenance is required. Periodic inspection is
recommended to insure proper functioning. It is recommended that a copy of the Calibration
Checklist (section 5.3.) be filled out whenever the PRM-470B is put into service after repair or
recalibration.
5.2. ELECTRONIC CALIBRATION PROCEDURE
The following is a list of the equipment required for the electronic calibration procedure. These
steps should only be performed by qualified service personnel.
• Oscilloscope
• Tweaker or small slotted screwdriver
• Digital voltmeter
• 5-10µCi
137
Cs test source
5.2.1. Gain (HV) Adjustment:
Refer to "HHV-454 High Voltage Power Supply Schematic Diagram" Drawing 6 and "HHV-454
High Voltage Power Supply Component Designator" Drawing 7 (Appendix F.).
Place
137
Cs source on the detector. Adjust the high voltage pot which is found in the copper
shielded H.V. board until a 3.2 volt pulse amplitude is obtained. Refer to "Typical Pulse Profile"
Drawing 10 (Appendix F.).