GENERAL PURPOSE
THERMAL MASS FLOWMETER
MODEL 4140/4143
OPERATION AND SERVICE MANUAL
P/N 1980383, REVISION G
FEBRUARY 2016

Questions and Answers

  • F
    Frank RussellAug 2, 2025
    What to do if TSI Incorporated 4140 Measuring Instruments flow readings fluctuate greatly?
    • P
      Paula GeorgeAug 2, 2025
      If you observe significant fluctuations in flow readings, it could be due to actual flow variations. To address this, try improving the inlet conditions or increasing the display averaging time.
  • T
    Thomas EverettAug 9, 2025
    Why does my TSI Incorporated 4140 Measuring Instruments temperature read high at low flows?
    • S
      Susan BryanAug 9, 2025
      A high temperature reading at low or zero flows can occur because the temperature sensor is being heated by the flow sensor. Once the flow rate through the flowmeter exceeds 1 Std L/min, the temperature value will accurately reflect the actual gas temperature.
  • A
    Angie BryantAug 17, 2025
    What to do if my TSI Incorporated 4140 display shows flows over-range?
    • R
      Ronald SanchezAug 17, 2025
      If the display indicates flows are over-range even when no flow is passing through the flow tube, the sensor may be damaged. In this case, return the flowmeter to TSI for service.
  • K
    Kimberly BranchAug 25, 2025
    How to fix no display on TSI Incorporated 4140?
    • A
      Anne WatsonAug 26, 2025
      If there is no display, ensure that the unit is switched on. Also, verify that the AC adapter is properly plugged in or check the power source connected to the mini-DIN connector.

Summary

Unpacking and Parts Identification

Setting-Up

Supplying Power

Details powering the flowmeter via AC adapter or mini-DIN.

Connecting Filter and Flow Tubes

Instructions for connecting filters and flow tubes to the flowmeter.

Configuration Software

Information on using TSI setup software for parameter changes.

Operation

Flow Rate Measurement

How to obtain flow rate data via display, RS232, or analog output.

Maintenance

Flow Sensor

How to inspect and clean the flow sensor using compressed air.

Troubleshooting

Specifications*

Flow Measurement

Details measurement range, accuracy, and resolution for flow.

TSI Incorporated 4140 Specifications

General IconGeneral
Weight1.3 lb (0.59 kg)
Operating Temperature32 to 100 °F (0 to 38 °C)
Accuracy±10%
Operating Humidity0 to 95% non-condensing
Power SupplyAC adapter
Battery Life> 8 hours (alkaline batteries)