ENERGY AND COMFORT

Table of Contents

Questions and Answers

  • T
    thomasryanSep 14, 2025
    What does Error Code 14 mean on TSI Incorporated Measuring Instruments?
    • R
      Richard EricksonSep 14, 2025
      Error Code 14 on your TSI Incorporated Measuring Instruments device indicates an accessory temperature/humidity probe checksum error for humidity. Perform user calibration or return to factory.
  • M
    Michael MaddoxAug 25, 2025
    What to do if TSI Incorporated Measuring Instruments displays flashing “888888”?
    • J
      Jennifer NealAug 25, 2025
      If your TSI Incorporated Measuring Instruments device is displaying flashing “888888”, it could be due to: * Dirty battery contacts: Clean the battery contacts. * The pressure is too high to measure: Check the specifications page for the maximum pressure. * The temperature is too high to measure: Consult the specifications page for the maximum temperature.
  • R
    Ray WashingtonAug 26, 2025
    What to do if TSI Incorporated Measuring Instruments shows Error Code 43?
    • D
      Dustin MoyerAug 26, 2025
      If your TSI Incorporated Measuring Instruments device displays Error Code 43, it indicates that the RS232 data could not be sent because the receiving device was busy for more than 10 seconds. To resolve this: * Ensure that the printer or computer is ready to receive data. * Check the RS232 interface cable.
  • M
    marycooperAug 26, 2025
    What does Error Code 39,40, 41,42 mean on TSI Incorporated Measuring Instruments?
    • C
      Carla JohnsonAug 26, 2025
      Error Codes 39, 40, 41, and 42 on your TSI Incorporated Measuring Instruments device indicate an error in the logged data. If you are using an accessory temperature/humidity probe, ensure it is securely connected. You may also need to erase the logged data.
  • D
    David GonzalezSep 15, 2025
    What does Error Code 15 mean on TSI Incorporated Measuring Instruments?
    • J
      Janice GarciaSep 15, 2025
      Error Code 15 on your TSI Incorporated Measuring Instruments device indicates an accessory temperature/humidity probe checksum error for temperature. Perform user calibration or return to factory.
  • W
    Whitney GilbertSep 23, 2025
    What does Error Code 74 mean on TSI Incorporated Measuring Instruments?
    • B
      burnettrhondaSep 23, 2025
      Error Code 74 on your TSI Incorporated Measuring Instruments device indicates that the pressure voltage for user pressure calibration is out of range. Re-do user calibration.
  • M
    Miss Stephanie Adams MDSep 7, 2025
    What does Error Code 59 mean on TSI Incorporated Measuring Instruments?
    • J
      Jennifer HarrisonSep 7, 2025
      Error Code 59 on your TSI Incorporated Measuring Instruments device indicates that the slope factor for user temperature sensor calibration is out of range. You should re-do user calibration, set temperature sensor calibration to FACT, or return to factory.
  • G
    Gregory PowellSep 8, 2025
    What does Error Code 60 mean on TSI Incorporated Measuring Instruments?
    • N
      Nathan HernandezSep 8, 2025
      Error Code 60 on your TSI Incorporated Measuring Instruments device indicates that the humidity sensor voltage at 20% RH is out of range. Perform user calibration or return to factory.
  • T
    Tracy FoleySep 2, 2025
    What does Error Code 30 mean on TSI Incorporated Measuring Instruments?
    • T
      Thomas MartinezSep 2, 2025
      Error Code 30 on your TSI Incorporated Measuring Instruments device indicates that the pressure calibration voltage at 1 in. H2O must be less than the pressure calibration voltage at 5 in. H2O. Make sure calibration pressures are correct.
  • J
    Jeffrey SandovalAug 31, 2025
    What does Error Code 10 mean on TSI Incorporated Measuring Instruments?
    • M
      Michael DouglasAug 31, 2025
      Error Code 10 on your TSI Incorporated Measuring Instruments device indicates that the AC adapter voltage is too low. Use the correct AC adapter.

TSI Incorporated 8710 DP-CALC Specifications

General IconGeneral
Measurement TypeDifferential Pressure
Accuracy±0.1 Pa
Pressure UnitsPa, inH2O, mmH2O, psi, mbar
DisplayLCD
Power Supply4 AA batteries or AC adapter
Operating Temperature0°C to 50°C
Storage Temperature-20°C to 60°C
WarrantyTwo years
Temperature Range0 to 50 °C (32 to 122 °F)