28 Model 8710 DP-CALC
TM
Micromanometer and Model 8375 ACCUBALANCE
®
Air Balancing Tool
4. Press X until the display shows SOURCE U-F for the desired measurement.
5. Press
. The display shows SOURCE and whether USEr or FACt is selected for that measurement.
6. Press S or T to select USEr or FACt.
7. Press
to save the selection.
Notes: For the accessory temperature/humidity probe it is not possible to mix user and factory
calibrations for the different measurements. For example, USEr cannot be selected for
temperature with FACt selected for humidity.
The hood flow calibration factor is applied when the offset factor is changed. There is no
option to select USEr or FACt calibration for hood flow measurement. Set value to 1.000 to
restore factory calibration.
Connect and Download Data to an Optional Printer
To connect a printer, use the RS232 printer cable supplied with the optional printer. Note that the printer
interface cable is different from the computer interface cable.
The printer must be set to the same baud rate as the instrument. To change the baud rate of the printer,
refer to the printer’s operations manual. Always turn the instrument on before the printer. If the printer
prints question marks (??????), asterisks (******), or random characters, reset it by turning it off and then
on again.
Figure 12 illustrates the printout information for a single sample.
---------------------------
MODEL: 8710 DP-Calc
SERIAL: 0404004
---------------------------
TEST ID: 2
Sample: 1
02/16/04
15:14:31 1.00 CFM
Shape=Round
Dia = 4.0 in. ø
Entered Temp= 74.4 °F
Baro. Press= 29.16 in.Hg
15:14:31 28.5 %rh
Figure 12: Sample Print Out
The Model 8710
allows printing of the entire memory, all Samples stored within a particular Test ID, or
an individual Sample.
To print the entire memory, press and hold the
key. This starts a countdown from 3 to 0 on the
display. When the display shows “0”, release the button. If you release the key at any time other than 0
during the countdown, nothing will print. The display will flash “Send dAtA” while printing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Incorporated 8710 DP-CALC and is the answer not in the manual?

TSI Incorporated 8710 DP-CALC Specifications

General IconGeneral
BrandTSI Incorporated
Model8710 DP-CALC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals