Chapter 7: Maintenance and Troubleshooting 49
Table 6: Error Codes
Error Code Possible Causes Corrective Action
10 The AC adapter voltage is too low.
The NiMH batteries will not recharge fully
with low AC adapter voltage.
Use correct AC adapter.
11 The AC adapter voltage is too high.
The NiMH batteries will not fast-charge
with high AC adapter voltage.
Use correct AC adapter.
12 Model number checksum error. Return to factory.
13 Serial number checksum error. Return to factory.
14 Accessory temperature/humidity probe
checksum error for humidity.
Perform user calibration or return to
factory.
15 Accessory temperature/humidity probe
checksum error for temperature.
Perform user calibration or return to
factory.
16 Accessory temperature probe calibration
checksum error.
Perform user calibration or return to
factory.
18 Pressure sensor calibration checksum error. Perform user calibration or return to
factory.
19 A to D converter calibration checksum
error.
Return to factory.
20 Barometric pressure sensor calibration
checksum error.
Return to factory.
30 Pressure calibration voltage at 1 in. H
2
O
must be less than pressure calibration
voltage at 5 in. H
2
O.
Make sure calibration pressures are
correct.
31 Pressure calibration voltage at 5 in. H
2
O
must be less than pressure calibration
voltage at 15 in. H
2
O.
Make sure calibration pressures are
correct.
32 While performing user calibration of
accessory temperature/humidity probe, the
correct probe was not found.
The accessory temperature/humidity
probe was either not plugged in or is
faulty.
33 While performing user calibration of
accessory temperature sensor, the
accessory temperature/humidity probe was
found instead.
Plug in the selected probe.
34 Barometric pressure measured was <15.00
in. Hg.
The barometric pressure is too low to
take a reading.
35 Barometric pressure measured was >40.00
in. Hg.
The barometric pressure is too high to
take a reading
36 Flap closed pressure was < flap open
pressure, or flap open versus flap closed
pressure ratio exceeds allowed limits.
Take data again.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Incorporated 8710 DP-CALC and is the answer not in the manual?

TSI Incorporated 8710 DP-CALC Specifications

General IconGeneral
BrandTSI Incorporated
Model8710 DP-CALC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals