4-4 Model 3786 Ultrafine Water-based Condensation Particle Counter
Communications Parameters
All serial communications with the UWCPC are accomplished using
the following communications parameters:
Baud Rate..........115,200
Bits/Character..............8
Stop bits .......................1
Parity ......................None
All data communication is preformed through ASCII-based
character codes.
All multi-field responses are comma separated values (CSV).
All input commands and output responses are terminated with a
carriage-return (CR).
All input Line-feeds (LF) codes are ignored.
Terminal Communications
Once the Serial or USB connection is made between the UWCPC
and a host computer, a terminal program may be used to
communicate with the instrument. Terminal emulation software
that can be used includes:
• Tera Term, a free software terminal emulator for MS-Windows
available at:
http://hp.vector.co.jp/authors/VA002416/teraterm.html
• HyperTerminal
™
which is included with most MS-Windows
operating systems.
• The programs, minicom and xminicom, are included with most
Linux distributions.
™