2-36 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
Installing Aerosol Instrument Manager Software
If the classifier will be used with the TSI Aerosol Instrument Manager
software, install the software on your PC. Aerosol Instrument Manager
software, versions 10.0 and above are compatible with the 3082 and 3938.
For installation requirements and install procedure, consult Chapter 2 of
the Aerosol Instrument Manager Software User’s Manual (P/N 1930038)
Revision K and above.
Connecting the Classifier to the Computer
The classifier can be connected to a computer (PC) using either USB or
Ethernet.
USB Communication
If using a USB connection, connect the supplied USB cable to the USB
Type B connection on the back of the classifier. Connect the other end of
the cable to the PC.
C a u t i o n
If the computer enters standby mode (may also be called sleep or
hibernate mode) when connected to the classifier, the instrument may
become unresponsive. If this occurs, the classifier must be power
cycled to recover. It is recommended that you disable standby mode on
your PC to avoid this.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 3938 and is the answer not in the manual?

TSI Instruments 3938 Specifications

General IconGeneral
BrandTSI Instruments
Model3938
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals