EasyManuals Logo

TSI Instruments 4000 Series User Manual

TSI Instruments 4000 Series
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
42
be used in the integral. If an end-trigger is set (set with
SETx
±nnn.nn), the acquisition will stop either when the end-trigger
condition is detected or when nnnn samples has been acquired,
whichever comes first. After the command is finished, a termination
sequence is sent to signal the end of the transfer.
Example 4)
DAFxx0005 Request 5 samples of flow in comma
delimited ASCII format (assume begin-trigger is not
set).
Flowmeter returns the acknowledgement “OK” <CR>
<LF> followed by the flow data and termination
sequence 1.10,1.20,1.25,1.23,1.20,1.10 <CR> <LF>
Example 5) DCFTxx0003 Request 3 samples of flow and
temperature in comma / <CR> delimited ASCII format
(assume begin trigger is not set).
Flowmeter returns the acknowledgement “OK” <CR>
<LF> followed by (flow) comma (temperature) <CR>
<LF> (flow) comma (temperature) <CR> <LF> (flow)
comma (temperature) <CR> <LF>
After importation into a spreadsheet it might look like
this:

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 4000 Series and is the answer not in the manual?

TSI Instruments 4000 Series Specifications

General IconGeneral
BrandTSI Instruments
Model4000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals