IDC [<string>]
Sets the main display to dc Amps and optionally sets the range; if no range is specified the
display defaults to mA autorange.
The following <strings> can be used to set the range:
<1MA>, <100MA>, <10A>
IAC [<string>]
Sets the main display to ac Amps and optionally sets the range; if no range is specified the
display defaults to mA autorange.
The following <strings> can be used to set the range:
<1MA>, <100MA>, <10A>
IACDC [<string>]
Sets the main display to ac+dc Amps and optionally sets the range; if no range is specified the
display defaults to autorange.
The following <strings> can be used to set the range:
<1MA>, <100MA>, <10A>.
OHMS [<string>]
Sets the main display to Ohms and optionally sets the range; if no range is specified the display
defaults to autorange.
The following <strings> can be used to set the range:
<100>, <1000>, <10K>, <100K>, <1000K>, <10M>, <20M>
CONT
Sets the main display to continuity test.
DIODE
Sets the main display to diode test.
CAP [<string>]
Sets the main display to Capacitance and optionally sets the range; if no range is specified the
display defaults to autorange.
The following <strings> can be used to set the range:
<10NF>, <100NF>, <1UF>, <10UF>, <100UF>
FREQ [<string>]
Sets the main display to Frequency and optionally sets the range; if no range is specified the
display defaults to autorange.
The following <strings> can be used to set the range:
<100HZ>, <1000HZ>, <10KHZ>, <100KHZ>
AUTO
Sets the main display to autorange, except for Continuity or Diode check.
MAN
Sets the main display to manual, i.e. the current range is ‘locked’.