AXB?
Returns the result of the Ax+b scaling function. The syntax of the response is:
<ASCII data><rmt>
where <ASCII data> is a 10-character string in the same format as the first (value) field of the
standard READ? response, see that command for details. If Ax + b is not running a zero value will
be returned.
WATTS [<nrf>]
Selects the Watts function and optionally sets the reference impedance; if no impedance is
specified the existing value is used. Vac or Vdc must first have been specified in the main display;
Watts shows in the secondary display.
WATTS?
Returns the result of the Watts calculation. The syntax of the response is:
<ASCII data><rmt>
where <ASCII data> is an 18-character string divided into two fields (value followed by units)
exactly as for the standard READ? response, see that command section for details. If the Watts
function is not running a zero value will be returned.
VA
Sets the VA function; Vac or Vdc must first have been selected in the main display and the meter
must be connected for both voltage and current measurement. VA shows in the secondary
display.
VA?
Returns the result of the VA calculation. The syntax of the response is:
<ASCII data><rmt>
where <ASCII data> is an 18-character string divided into two fields (value followed by units)
exactly as for the standard READ? response, see that command section for details. If the VA
function is not running a zero value will be returned.
CANCEL
Cancels the second level modifier currently running.
Data Logging Commands
LOGON [<nrf>]
Sets Data Logger running and optionally sets the internal logger interval timer; when no interval is
specified the existing value is used.
Setting the interval to 0000 disables the timer but still allows the logger to be triggered manually
(from the keyboard) or by remote command (TRIG), see below; triggering by remote contact
closure is not possible during remote control operation.
LOG?
Returns the contents of all valid locations in the data logger store, starting at location 001.
The syntax of the response is:
<ASCII data><rmt>
where <ASCII data> consists of up to 100 complete results, separated by commas, each result
comprising three fields. The first field is the reading number, consisting of 3 digits followed by