V (for diode test) dB
W VA %
Examples:
101.23e-3 V DC (101.23mV)
-10.001e00 V DC (-10.001V)
00.123e00 V AC+DC (0.123V)
100.01e03 Hz (100.01kHz)
01.010e-6 F (1.01µF)
READ2?
Returns the next reading from the secondary display immediately after the command has been
parsed. The syntax is as for READ? described above. If the secondary display is currently
showing the range of the main display the response will be RANGE.
EVERY
Sends every reading from the main display until stopped by STOP or any other command. The
syntax of each reading is:
<ASCII data><rmt>
where <ASCII data> is an 18-character string divided into two fields (value followed by units)
exactly as for the standard READ? response, see that command section for details.
STOP
Cancels the EVERY command.
RST
Resets the meter to the factory default settings except for the Address and Baud rate of the
RS232 interface itself.
Main Display Commands
VDC [<string>]
Sets the main display to dc Volts and optionally sets the range; if no range is specified the display
defaults to autorange.
The following <strings> can be used to set the range:
<100MV>, <1000MV>, <10V>, <100V>, <1000V>
VAC [<string>]
Sets the main display to ac Volts and optionally sets the range; if no range is specified the display
defaults to autorange.
The following <strings> can be used to set the range:
<100MV>, <1000MV>, <10V>, <100V>, <750V>
VACDC [<string>]
Sets the main display to ac+dc Volts and optionally sets the range; if no range is specified the
display defaults to autorange.
The following <strings> can be used to set the range:
<100MV>, <1000MV>, <10V>, <100V>, <750V>