Why does the OVERLOAD indicator light up when I switch the high voltage ON with my Ulvac HPS-600N?
- JjwatkinsNov 21, 2025
If the high voltage is switched to ON and the OVERLOAD indicator lights up, immediately switching the high voltage to OFF, it could be due to several reasons: * A ground fault in the high voltage line of the EB gun: Inspect the high voltage line, ensuring the OUTPUT cable of the EB power supply is disconnected, and use a 1,000V megohmmeter to check the insulation. * Emitter assembly insulation failure: Refer to the EB gun's operations manual to inspect the emitter assembly. * High voltage introduction terminal insulation failure: Examine the isolation insulator, replacing it if there’s significant dirt from deposited film. * Glow electric discharge: Ensure the pressure is set to 5 x 10-3 Pa or less. * Power supply breakdown: Check if HV ON works with no load. Note: This is dange...