IM 50.560CA UA (B/4-03) 19
DEPOLOX
®
3 PLUS RESIDUAL ANALYZER
1278 349105 6 11 12
L1 L2/N PE
37 38 39 32 3031 33 25 26 2934 35 36 27 28
AB-+
CAUTION: CONSIDER
TENSION MODULE !
POWER SENSOR
FREE CHLORINE
TEMPER-
ATURE
CELL
DIGITAL
IN
mA-
OUT
CHLORINE
RS485-BUS
BN
GN
RD
WH
BU
BK
123
YE
GN
BN
WH
BU
RD
132
L1 L2/N PE
PE
PT100
DEPOLOX 3 PLUS
CELL
BARE ELECTRODE
IF SAMPLE WATER STOP
CONTACT IS OPEN
e.g. FLOW CONTROL
(OPTION)
TO LINE RECORDER, PLC
OR SCADA-SYSTEM
3 COND.
YE = YELLOW
BN = BROWN
GN = GREEN
BU = BLUE
WH = WHITE
RD = RED
GY = GREY
BK = BLACK
COLOR FROM CABLE
115V AC
230V AC
50/60 Hz
LINE VOLTAGE
POWER
CNT
REF
WRK
GY
YE
GY
14 AWG
7 COND./SHD
32 AWG 32 AWG
2 COND.
20 AWG
2 COND./SHD
20 AWG
2 COND./SHD
DEPOLOX 3 PLUS WITH BARE ELECTRODE SENSOR KIT - WIRING
50.560.155.010
ISSUE 1 1-03
NOTE: FIELD WIRING (NOT BY USF/W&T) MUST CONFORM TO LOCAL ELECTRICAL CODES.