IM 50.560CA UA (B/4-03)
DEPOLOX
®
3 PLUS RESIDUAL ANALYZER
44
Span calibration
a. After zero calibration wait at least two minutes. Then take a
sample at the measuring cell. Determine the residual value of
this sample.
b. Press the key until the menu “DIS span” is reached. Press the
E key to open the menu.
c. Press keys or until the displayed value agrees with the manu-
ally measured value.
d. Store the value using the E key. The chlorine measuring cell is
now calibrated. In case of an error message, refer to paragraph
4.1.1, Errors.
e. Return to the basic display by pressing the key twice.
3.2.2 Membrane Sensor Calibration (All Membrane Types)
a. Take a sample at the measuring cell. Determine the residual value of
this sample.
b. Starting from the basic display, press the F key until the “Calibra-
tion” menu is displayed.
c. Press the key until the menu “DIS span” is reached. Press the E
key to open the menu.
d. Press keys or until the displayed value agrees with the manu-
ally measured value.
e. Store the value using the E key. The disinfectant measuring cell is
now calibrated. In case of an error message refer to paragraph 4.1.1,
Errors.
f. Return to the basic display by pressing the key twice.