8UV Smart D25 | Instructions for use | Revision 9 EN
Claim Micro-organism Sort and type Log reduction
Bactericidal
Staphylococcus aureus Gram-positive bacteria, vegetative ≥ 5,0
Pseudomonas aeruginosa Gram-positive bacteria, vegetative ≥ 5,0
Enterococcus hirae Gram-positive bacteria ≥ 5,0
Escherichia coli Gram-negative bacteria, Enterobacteriaceae ≥ 4,0
Fungicidal
Candida albicans Fungus, vegetative yeast ≥ 4,0
Aspergillus brasiliensis Fungus, spore ≥ 4,0
Yeasticidal
Candida albicans Fungus, vegetative yeast ≥ 4,0
Candida auris Fungus, yeast ≥ 4,0
Sporicidal Bacillus subtilis Gram-positive bacteria, endospore ≥ 4,0
Mycobactericidal
Mycobacterium avium Gram-positive mycobacteria, vegetative ≥ 4,0
Mycobacterium terrae Gram-positive mycobacteria, vegetative ≥ 4,0
Tuberculocidal Mycobacterium terrae Gram-positive mycobacteria, vegetative ≥ 4,0
Virucidal
Adenovirus type 5, strain Adenoid 75 Adenoviridae ≥ 4,0
Bovine coronavirus type 1 Coronaviridae ≥ 4,0
Poliovirus type 1; LSc-2ab (picornavirus) Enterovirus ≥ 4,0
Murine norovirus, strain s99 Berlin Caliciviridae ≥ 4,0
Polyomavirus SV40, strain 777 Incertae sedis ≥ 4,0
Claimed performance
The required reduction of micro-organisms is based on the EN 14885:2018 to
compare it to the state of the art of chemical disinfectants and antiseptics. The
following reductions are determined and claimed: