6. TROUBLE SHOOTING
Syndrome Scale cannot be turned on
Check: Is the scale charged?
Action: Recharge the scale for at least 8
hours before first time use or plug
in the power adaptor before power
on the scale.
Check Is the power adaptor inserted
properly into both the electricity
outlet and the DC inlet of scale.
Action: Secure both ends of the power adaptor
and try again.
Syndrome Scale turned off automatically
Check Is Auto Power Off Function employed?
Action Refer to 4.2.2 to disable Auto Power
Off Function.
Check Is the LO-BAT INDICATOR on?
Action Apply the power adaptor.
Syndrome Rated capacity cannot be reached
Check Is the TARE INDICATOR on?
Action: Turn the scale off. Remove all loads
from platter and turn on again.
Check Is there anything obstructing the
platter?
Action Remove all obstacles.
Syndrome Blank display with only "kg" or "lb"
Check Is the load applied to platter excess
the rated capacity of scale?
Action Remove all loads from platter and
try again.
Syndrome Scale displays
-----
Check Is the platter inserted to the scale
correctly?
Action Insert the platter to scale
correctly.
Check Is the stainless steel platter cover
placed on the platter?
Action Place the stainless steel platter
cover onto the platter.