35
The faults, indicated in the table below, are listed in order of priority. The faults on top of the list have
the highest priority. This means that faults with a lower priority can be overwritten by faults of a higher
priority.
At power-on the system is explicitly tested for failures indicated by display messages 1, 2, 3, 4, 5, 6
and 7. During normal operation these tests (except 2) run at the background.
下表中指出的故障系按优先顺序排列。表中上方的故障有最高优先权。这意味着拥有高优先的故障可
以覆盖低优先问题。当接通电源后,系统将明确地检测是否存在故障,此时显示器依次显示信息1, 2,
3, 4, 5, 6和7。在正常操作情况下,这些测试(2 除外)将在本底进行。
or or
Followed by 接下来:
Display
message
显示信息
iscosity
output [mA]
黏度输出
Temperature
output [mA]
温度输出
Recovery action
(attempt)
恢复动作(尝试)
Fault description
故障描述
0
0
Delay & retry
15 V power failure. Supply drops below 13 V
15伏电源故障。供电电源在13伏以下
0
Actual
Delay & retry
Pendulum not oscillating 钟摆无
振动
4
Actual
Delay & retry
Oscillation error 振动误差
4
4
20
4
None
None
Temperature out of range (overflow)
温度超出量程(溢出)
Temperature out of range (underflow)
温度超出量程(下溢)
20
4
Actual
Actual
None
None
Viscosity out of range (overflow)
黏度超出量程(溢出)
Viscosity out of range (underflow)
年度超出量程(下溢)
Actual
Open loop
Open loop
Actual
None
None
Temperature output open loop 温度输出开环回路
Viscosity output open loop 黏度输出开环回路
Not all segments
working after
power on
Actual
Actual
Not applicable
7 - segment display defective 7段显示欠缺
Any other figure or letter indicates an error in the electronics or software.