EasyManua.ls Logo

Vaisala HMP60 - Repair Maintenance; Changing the INTERCAP Sensor (HMP60); Figure 17 Changing the Sensor

Vaisala HMP60
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide ______________________________________________________________________
54 __________________________________________________________________ M211060EN-C
Repair Maintenance
Changing the INTERCAP® Sensor
(HMP60)
This procedure restores the humidity measurement accuracy of the
HMP60. No adjustment after the sensor change is needed.
To perform this procedure, you need a new INTERCAP® sensor. It is
also recommended that you replace the filter with a new one. For order
codes, see section Options and Accessories on page 62.
1. Rem
ove the filter from the probe by turning it counter-clockwise.
2. Pull out the old INTERCAP® sensor (the taller of the two sensors
inside the probe) and insert a new one. Refer to Figure 17 on
page 54. Handle the new sensor by the plastic frame. DO NOT
TOUC
H THE SENSOR PLATE.
3. Attach a new filter on the probe.
0912-097
Figure 17 Changing the Sensor

Table of Contents

Related product manuals