M211060EN-J
User Guide
Vaisala Humidity and Temperature Probes
HMP60 and HMP110 Series

Table of Contents

Questions and Answers

  • K
    Kimberly Burke MDAug 19, 2025
    How to troubleshoot Vaisala Accessories measurement not working?
    • M
      Mark GoldenAug 20, 2025
      If the Vaisala Accessories measurement is not working, start by opening the filter and checking the sensors for damage or if any are missing. Also, verify the probe's supply voltage and check the output mode of the probe using a serial line.
  • T
    Tiffany MccormickAug 21, 2025
    What to do if Vaisala Accessories probe is not responding to serial commands?
    • K
      Kathleen ValdezAug 21, 2025
      If your Vaisala Accessories probe isn't responding to serial commands, try disconnecting the USB cable and retry. If that doesn't work, use the '??' command in case the probe is in POLL mode, and open the line using the OPEN command if necessary. Also, check your serial line settings and cable connections.
  • J
    Jennifer MartinAug 23, 2025
    Why Vaisala Accessories humidity measurement appears to be wrong?
    • M
      Marie LewisAug 25, 2025
      If the humidity measurement of your Vaisala Accessories appears to be wrong, check the probe and sensor for condensation. Use the L command to check the currently applied calibration correction. For HMP110, calibrate and adjust the probe. For HMP60, calibrate the probe and change the sensor if necessary.
  • P
    Phillip HoodAug 26, 2025
    What to do if Vaisala Accessories shows a voltage error?
    • M
      Madison RayAug 26, 2025
      If you see a voltage error active on your Vaisala Accessories, it means the operating voltage is out of range. Correct the voltage and then reset the probe.
  • T
    Tony RobertsAug 27, 2025
    Why is the humidity measurement wrong on my Vaisala Tent?
    • M
      Michael TerryAug 28, 2025
      If the humidity measurement of your Vaisala Tent seems incorrect, first check the probe and sensor for any condensation. You can also use the L command to check the currently applied calibration correction. Another solution is to calibrate and adjust the probe (HMP110). If you are using HMP60, calibrate the probe and change the sensor if necessary.
  • D
    Deanna SchwartzAug 30, 2025
    What to do if the measurement is not working on my Vaisala Tent?
    • A
      Anna KingAug 30, 2025
      If the measurement is not working on your Vaisala Tent, start by opening the filter and checking the sensors for any damage or if they are missing. Also, check the supply voltage of the probe and the output mode of the probe using serial line.
  • N
    Nicholas DuncanAug 31, 2025
    How to fix a Vaisala Tent probe that is not responding to serial commands?
    • W
      williamhunterAug 31, 2025
      If the probe of your Vaisala Tent isn't responding to serial commands, try disconnecting the USB cable from the probe and retry. Also, try the ?? command in case the probe is in POLL mode, open the line using OPEN command if necessary. Ensure your serial line settings and cable connections are correct.
  • N
    Nancy HinesSep 2, 2025
    What causes a voltage error on my Vaisala Tent and how to fix it?
    • M
      Michelle WoodSep 2, 2025
      If your Vaisala Tent displays a voltage error, it means the operating voltage is out of range. Correct the voltage and then reset the probe.
  • C
    Christina YangSep 4, 2025
    What does checksum error mean on Vaisala Tent?
    • J
      john70Sep 4, 2025
      If a checksum error occurs on your Vaisala Tent, it may indicate an internal error. Try restoring the probe to factory settings using the FRESTORE command and check again. If the error persists, no answer.
  • V
    victor96Sep 6, 2025
    What to do if the supply voltage is out of range for Vaisala HMP110?
    • G
      Gary BowenSep 6, 2025
      If the supply voltage for your Vaisala Measuring Instruments is too low, check and correct the power supply and wiring. If there's an internal transmitter failure, power-cycle the device and, if needed, restore the factory settings. If the error continues, contact Vaisala.

Summary

About this Document

Version Information

Details document versions and update history for HMP60/HMP110 probes.

Related Manuals

Lists other Vaisala manuals relevant to the HMP60/HMP110 series probes.

Product Overview

Introduction to HMP60 and HMP110 Series

Introduces HMP60/HMP110 probes and their applications.

Basic Features and Options

Covers analog/digital outputs, RS-485, size, power, body material, and accessories.

Installation

Mounting HMP60, HMP110, and HMP110 T Probes

Instructions for mounting HMP60/HMP110/HMP110T probes using M12 thread or accessories.

Wiring

Provides pin-out diagrams and grounding methods for HMP60/HMP110 series probes.

Communication and Software

Using Multiple Devices over RS-485

Discusses connecting multiple HMP60/HMP110 probes using RS-485.

Connecting to Insight Software

Steps for connecting HMP60/HMP110 probes to Vaisala Insight software.

Serial Line Communication

Connecting to Serial Interface

Instructions for connecting to the probe's serial interface using USB cable.

Installing the USB Cable Driver

Guide for installing the Vaisala USB driver on a PC.

List of Serial Commands

Lists available serial commands for probe configuration and operation.

Configuration and Calibration

Configuring Serial Line Operation

Commands to set serial line settings like baud rate, parity, and interface mode.

Calibration Commands

Commands for calibrating humidity and temperature measurements.

Calibrate Humidity Measurement

Using CRH command for 1-point or 2-point RH calibration.

Calibrate Temperature Measurement

Using CT command for 1-point or 2-point temperature calibration.

Maintenance and Repair

Periodic Maintenance

Covers yearly calibration and sensor replacement for accuracy.

Adjustment Procedure using Serial Line

Serial line adjustment procedures for HMP110/HMP113 RH measurement.

Repair Maintenance

Procedures for changing sensors on HMP60/HMP63 and HMP110/HMP113.

Troubleshooting

Solving Typical Problems

Troubleshooting common issues using serial interface and ERRS command.

Error Messages in Insight Software

Lists common errors in Vaisala Insight software and their causes/solutions.

Technical Data

HMP60 Specifications

Detailed measurement performance, operating environment, and I/O specs for HMP60.

HMP110 Specifications

Detailed measurement performance, operating environment, and I/O specs for HMP110.

Probe Dimensions

Provides dimensional drawings for HMP60/HMP110/HMP110T and HMP63/HMP113 probes.

Modbus Reference

Communication Settings

Details default Modbus serial communication settings.

Data Encoding

Explains 32-bit float/integer and 16-bit integer data encoding for Modbus.

Modbus Registers

Overview of Modbus registers, including measurement, configuration, and status.

Support and Warranty

Technical Support

Contact information and required details for technical support.

Maintenance and Calibration Services

Information on Vaisala's customer care and factory services.

Vaisala HMP110 series Specifications

General IconGeneral
BrandVaisala
ModelHMP110 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals