EasyManua.ls Logo

Vaisala HMP110 series - Technical Data; HMP60 Specifications

Vaisala HMP110 series
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9. Technical data
9.1 HMP60 specifications
Table 17 HMP60 measurement performance
Property Description/Value
Relative humidity
Measurement range 0 … 100 %RH
Typical accuracy:
at 0 … +40 °C (+32 … +140 °F) ±3 %RH (0 … 90 %RH)
±5 %RH (90 … 100 %RH)
at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F
and +104 … +140 °F)
±5 %RH (0 … 90 %RH)
±7 %RH (90 … 100 %RH)
Humidity sensor Vaisala INTERCAPâ
Temperature
Measurement range −40 … +60 °C (−40 … +140 °F)
Accuracy:
at +10 … +30 °C (+50 … +86 °F)
at −40 … +10 and +30 … +60 °C (−40 … +50 and
+86 … +140 °F)
±0.5 °C (±32.9 °F)
±0.6 °C (±33.08 °F)
Dew point temperature
Measurement range −40 … +60 °C (−40 … +140 °F)
Typical accuracy:
at +32 … +40 °C (0 … +140 °F)
when dew point depression
1)
< +15 °C (< +59 °F)
±2 °C (±35.6 °F)
at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F
and +104 … +140 °F) when dew point depression
< +10 °C (< +50 °F)
±3 °C (±37.4 °F)
Analog outputs
Accuracy at +20 °C (+68 °F) ±0.2 % of FS
Temperature dependence ±0.01 % of FS/°C (±0.006 % of FS/°F)
1) Dew point depression = ambient temperature − dew point.
HMP60 and HMP110 Series User Guide M211060EN-J
70

Table of Contents