EasyManua.ls Logo

Vaisala HUMICAP HMM170 - Status Registers; Table 21 Modbus Status Data Registers (Read-Only)

Vaisala HUMICAP HMM170
51 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Communication
1539 0602
hex
Parity, data, stop bits enum 0 = None, 8, 1
1 = None, 8, 2
2 = Even, 8, 1
3 = Even, 8, 2
4 = Odd, 8, 1
5 = Odd, 8, 2
(default: 1 = None, 8, 2)
1540 0603
hex
Response delay 16-bit integer Unit: ms
Range: 0 ... 1000
Functions
1285 0504
hex
Purge 16-bit function
status
When writing to
register:
1 = Activate purge
1542 0605
hex
Restart device 16-bit function
status
When writing to
register:
1 = Restart the device
A.4.3 Status registers
Table 21 Modbus status data registers (read-only)
Register
number
Address Register description Data
format
Note
513 0200
hex
Error status 16-bit
boolean
0000
hex
: One or more errors
active
0001
hex
: No errors
514 0201
hex
Online status 16-bit
boolean
0000
hex
: Output locked
0001
hex
: Online data
available
516 0203
hex
Error code 32-bit
signed
integer
See Table 22 (page 41).
0204
hex
518 0205
hex
Security hash 32-bit
signed
integer
Security hash changes when
any change is made to device
settings or adjustments.
0206
hex
HMM170 User Guide M212259EN-D
40

Table of Contents