USER'S GUIDE____________________________________________________________________
64 __________________________________________________________________ M210543EN-E
Example for PWD22:
CONF. PASSWORD (4 CHARS MAX)
UPDATE CONFIGURATION PARAMETERS
UNIT ID (2 CHAR) () 1
UPDATED
SET REFERENCE PARAMETERS
OFFSET ( 156.47) Y
OFFSET REFERENCE UPDATED
MODE UPDATED
ALARM LIMIT 1 ( 0) 1500
ALARM LIMIT 1 UPDATED
ALARM LIMIT 2 ( 0) 1000
ALARM LIMIT 2 UPDATED
ALARM LIMIT 3 ( 0) 500
ALARM LIMIT 3 UPDATED
RELAY CONTROL MODE
0 = 3*VIS, 1 = 2*VIS & HW ( 0) 1
RELAY ON DELAY ( 5)
RELAY OFF DELAY ( 5)
TRANSMITTER CONTAMINATION LIMITS
WARNING LIMIT ( 0.5)
ALARM LIMIT ( 3.0)
RECEIVER CONTAMINATION LIMITS
WARNING LIMIT ( 300)
ALARM LIMIT ( 600)
ANALOG OUTPUT MODE
0=LINEAR, 1=LN, 2=EXTI, 3=VGI, 4=VG2 ( 4)
ANALOG OUTPUT RANGE
MAX VISIBILITY ( 20000)
= mA ( 20.0)
MIN VISIBILITY ( 10)
= mA ( 4.0)
HOOD HEATERS USED 1=YES, 0=NO ( 0)
END OF CONFIGURATION
NOTE
The output illustrates the response to the command when using PWD22.
With PWD52 the maximum visibility is 35000 meters. The system
responses are otherwise identical when using PWD22 and PWD52.
The meaning of the questions asked by the system is described below.
Bold text is used to indicate user actions. The old settings are not
changed, if the user just presses ENTER as the answer.