16
Option KK: 4-Wire Kelvin Lead Set
Option "KK" is a heavy-duty 4-wire Kelvin cable set, 48-inches in length terminated in
"JAWS" clamps for measuring large motors, bushings, etc. Opening is 2".
Option MP-1: Kelvin Micro-Probes
Option "MP-1" is a 48-inch shielded Kelvin 4-wire cable set with a 1-ampere test current
capacity employing a set of Kelvin Micro-Probes. The probes are equipped with spring-
loaded stainless steel tips with 0.05" spacing.
Option MP-2: Kelvin Mini-Probes
Option "MP-2" is a 48-inch shielded 4-wire cable set equipped with Kelvin Mini-Probes
having spring-loaded stainless steel tips with 0.18" spacing.
Option MP-3: Kelvin Micro-Probes
Option "MP-1" is a 48" shielded lead set (dual banana), handheld pencil type, terminated in
spring loaded steel tips with .08" separation (1 amp max.). Special orders or modified
versions available upon request.
Option MP-4/MP-5: Surface Probes
These probes permit rapid, repeatable bonding testing on a variety of screened or flat
surfaces. Test current is evenly distributed through the probe base while sensing is
accomplished via a spring loaded center contact. The MP-4 target area is 1" in diameter.
The MP-5 target area is .4" in diameter.
Option MP-S: Single Probe Lead Set
Option "MP-S" is a 4-wire lead set terminated in single points. The 4-wire configuration is
maintained up to the point of the probe, eliminating most cable resistance effects. Option
MP-S may be used where a single probe tip is a must.
Option BBL: Banana-to-Banana Cable
Option "BBL" is a 48" shielded cable terminated on both ends in dual banana plugs. This
cable may be used for voltage and current connections to the ohmmeter.
Option C: Banana-to-Clip Cable
Option "C" is a 48" general purpose shielded lead set terminated on one end in dual banana
plugs and on the other end in red and black alligator clips.
Option SL-48: Low Thermal Leads
Option "SL-48" is a 48" shielded lead set terminated in gold-plated spade lugs. This lead set
is designed to eliminate problems caused by thermal EMF's and is rated for the maximum
output current of the instrument.