4150ATC Calibration Procedure
Voltage Zero Adjustment
1) Connect the voltage standard (2701C) at 0 volts to the VOLTAGE HI and LO terminals.
2) Select the 200 range and the STD mode.
3) Adjust the front panel ZERO potentiometer so that the display indicates 000.00.
Voltage Full-scale Adjustment
1) Set the 2701C to output 10mV. The ohmmeter uses negative voltages therefore the VOLTAGE
HI terminal should be negative with respect to the VOLTAGE LO terminal.
2) Adjust RV101 for a display indication of 100.00.
3) Remove all connections to the terminals.
Current Zero Adjustment
1) Using the 4-wire test leads, connect the input terminals of the ohmmeter to the 0.1 ohm standard
resistor and select the 200 range.
2) Adjust the front panel ZERO potentiometer for a display indication of 000.10 ohms ±1 digit.
20m Range Adjustment
1) Connect the 4150ATC to the 0.01 ohm standard resistor and select the 20m range.
2) Adjust R60 for a display indication of 10.000 ±2 digits.
200m Range Adjustment
1) Connect the ohmmeter to the 0.1 ohm standard resistor and select the 200m.
2) Adjust R64 for a display indication of 100.00 ±2 digits.
2 Range Adjustment
1) Connect the ohmmeter to the 1 ohm standard resistor and select the 2 range.
2) Adjust R66 for a display indication of 1.0000 ±2 digits.
20 Range Adjustment
1) Connect the ohmmeter to the 10 ohm standard resistor and select the 20 range.
2) Adjust R70 for a display indication of 10.000 ±2 digits.