20 Range Adjustment
1) Connect the ohmmeter to the 10 ohm standard resistor and select the 20 range.
2) Adjust R66 for a display indication of 10.000 ±2 digits.
200 Range Adjustment
1) Connect the ohmmeter to the 100 ohm standard resistor and select the 200 range.
2) Adjust R70 for a display indication of 100.00 ±2 digits.
2K Range Adjustment
1) Connect the ohmmeter to the 1Kohm standard resistor and select the 2K range.
2) Adjust R74 for a display indication of 1.0000 ±2 digits.
20K Range Adjustment
1) Connect the ohmmeter to the 10Kohm standard resistor and select the 20K range.
2) Adjust R76 for a display indication of 10.000 ±2 digits.
End calibration for 4100ATC.