User%guide%for%Cary%5000%absorption%spectrometer%
with%external%DRA%1800%attachment.%
(last&updated&03/17/2016)&
This"guide"is"for"use"of"the"Cary"5000"with"the"DRA"only.""For"use"without"the"DRA"as"a"
standard"transmission"spectrometer"see"the"User"guide"for"Cary"5000"in"absorption"
mode."
Importa n t%w a rn in gs !%
Do"not"unplug"or"plug"in"the"external"DRA"attachment"when"the"instrument"is"
on.""This"will"ruin"the"detectors"of"the"integrating"sphere"
Do"not"put"white"light"into"the"DRA"with"the"DRA"electrically"connected"to"the"
Cary."
Minimize"the"amount"of"time"that"the"integrating"sphere"is"open."Long-term"light"
exposure"hurt"the"detector"over"time."""
Wear"clean"gloves"while"using"the"instrument"
Do"not"use"liquid"or"powder"samples"in"the"integrating"sphere."
Quick%Start%for%using%DRA%
1. Turn"on"computer""
2. Check"lens"and"mirror"are"correct"for"your"measurement."
a. Install"Small"Sample"Kit"mirror"(M3"SSK)"or"regular"M3"mirror"if"needed"
b. If"using"SSK"install"correct"lens"L2."
3. Make"sure"cover"for"the"DRA"is"fully"closed"and"turn"Cary"on."
4. Click"on"the"Cary"icon"on"the"taskbar"
5. Wait"20"minutes"for"the"lamps"to"warm"up."
6. Select"Setup"
a. set"wavelength"range,"%T"or"%R"
b. Under"Options"tab:"set"slit"width"and"height,"SBW"to"2"nm,"Detector"and"
Grating"change"wavelengths"
c. Under"Baseline"tab:"chose"zero/baseline"
d. Under"Storage"tab:"Set"your"filename"
7. Make"sure"that"all"exits"of"the"integrating"sphere"are"blocked"with"full"reflectors."
8. Baseline%and"run"100%"T"scan,"then"block"the"sample"beam"at"the"entrance"to"the"
integrating"sphere"for"the"0%"T"scan.""If"using"an"aperture"hol der"for"reflection"
measurements,"please"see"the"“Aperture"kit"for"small"samples”"section"on"page"4."""
9. Mount"your"sample"
10. Run"Spectra"
11. When"done"remove"sample"and"remount"reflectors."
12. Turn"off"Cary"
Question and Answer IconNeed help?

Do you have a question about the Varian Cary 5000 and is the answer not in the manual?

Varian Cary 5000 Specifications

General IconGeneral
TypeUV-Vis-NIR Spectrophotometer
Wavelength Range175 - 3300 nm
Scanning SpeedUp to 2400 nm/min
Light SourceDeuterium and Tungsten-Halogen lamps
Detector TypePhotomultiplier Tube (UV-Vis), PbS (NIR)
Stray Light<0.05% at 220 nm
Baseline Flatness< +/- 0.0005 A (200 - 800 nm)
Baseline Stability< 0.0002 A/hour
Sample CompartmentStandard 10 mm cuvette holder

Summary

Important warnings!

Quick Start for using DRA

Introduction and Help

Turning the instrument on

Procedure for powering on the Cary instrument and software for initialisation.

Basics configuration and alignment

Small samples

Basics of measurement acquisition

Cary tab

Defines scan parameters like units, axes ranges, and scan controls.

Options tab under Setup

Controls Beam Mode, Slit Band Width, Slit Height, and detector/grating wavelengths.

Baseline tab

Enables selection of baseline correction types for scan runs.

Storage tab under Setup

Commands Menu

Graph menu

Trace Preferences

Allows selection and customization of traces displayed in graphs.

Graph Preferences

Enables changes to the overall appearance and layout of graphs.

Storing Data

Appendix A

How the Cary 5000/6000i Controls SBW and Energy

Explains control of Slit Band Width (SBW) and Energy level for different regions.

Appendix B

Acquiring, storing, and retrieving baseline files

Details the process of collecting, saving, and recalling baseline data.

Appendix C

Graph menu, Trace Preferences and Graph Preferences options

Details options within the Graph menu for trace and graph customization.