Introduction and Initial Operation
Page 6 User Manual Micro-GC Varian, Inc.
Preparing for Installation
Refer to the Pre-Installation Instructions (CP501289) for site
preparation and unpacking information.
GAS SUPPLY
Gas quality specifications are listed in the reference document “Pre-Installation
requirements Micro-GC CP501289”.
GAS SAMPLES
- Do not introduce liquid samples into your Micro-GC.
- Type of samples: non-condensing gas
- Samples other than non-condensing gases (wet, vapors, particles and
polymers) must be filtered in advance.
- Sample conditions: Non condensing gas between 0 and 110C.
- Sample pressure between 0 and 100 kPa (1 bar, 15 psi)
- Outlet of sample container must fit to a stainless steel capillary of 1/16"
outside diameter, provided with a Swagelok
® female nut.