S-19-01-1214-UV101 V1.3 
Section 9: Accessories 
 
1.  Ultrasonic insert (optional) 
2.  AC Power Cord Set 
3.  Foot control switch 
4.  Water tubing with quick connector 
 
Section10: Techniques For Use 
 
10.1 Patient Positioning 
 
‧ For  optimal  access  to  both  the upper  and  lower 
arches,  the  backrest  of  the  chair  should  be 
adjusted  to  a  45∘   degree  angle.  This  assures 
patient comfort and Clinician visibility. 
‧ Have the patient turn his/her head to the right or 
left.  Also  position  chin  up  or  down  depending 
upon  the  quadrant  and  surface  being  treated. 
Evacuate irrigate using either a saliva ejector or 
High Volume Evacuator (HVE). 
 
10.2 Performing Ultrasonic Scaling 
Procedures 
‧ Note: Refer to the  Infection Control Information 
Booklet  supplied  with  your  system  for  general 
procedures  to  be  followed  at  the  beginning  of 
each day and between patients. 
‧ The  edges  of  ultrasonic  inserts  are  intentionally 
rounded  so  there  is  little  danger  of  tissue 
laceration  with  proper  ultrasonic  Scaling 
technique. Whenever the insert tip is placed in the 
patient’s Mouth, the lips, check and tongue should 
be  retracted  to  prevent  Accidental  prolonged 
contact with the activated tip. 
‧ Hold the empty Handpiece in an upright position. 
Activate the Foot Control until fluid exits. 
‧ Lubricate  the  rubber  O-ring  on  the  insert  with 
water before placing it into the handpiece. Fully 
seat insert  with  a  gentle push-twist motion.  DO 
NOT FORCE IT INTO PLACE. 
‧ Activate the System. Hold the handpiece over a 
sink or drain. Check spray temperature to verify 
fluid is reaching the working end of the insert tip. 
Adjust the water cooler irrigate. Control knob to 
ensure  adequate  flow  for  the  selected  Power 
setting.  Greater  flow  settings  provide  cooler 
irrigation. 
‧ It may be necessary to adjust  water flow larger 
under "Turbo" mode (Foot Control fully depressed) 
so adequate fluid will be available to cool tip and 
tooth interface. 
‧ In general, it is suggested a "feather-light-touch" 
be used both supra and subgingivally. The motion 
of  the  activated  tip  and  acoustic  effects  of  the 
irrigating  fluid,  in  most  cases,  is  adequate  to 
remove even the most tenacious calculus. 
‧ Periodically  check  the  ultrasonic  insert  for  wear 
with the Insert Efficiency Indicator. 
‧ The  use  of  a  saliva  ejector  or  High  Volume 
Evacuator  (HVE)  is  recommended  during  all 
procedures. 
‧ Set the System's Power Adjustment knob to the 
lowest power setting for the application and the 
selected insert. 
‧ If  water  leakage  found  in  handpiece,  replace 
sleeve or o-ring on handpiece cable assembly to 
eliminate. 
 
10.3 Patient Comfort Considerations 
Reasons for sensitivity 
‧ Incorrect tip placement. Point should be directed 
away from root surfaces. 
‧ Not keeping tip in motion on tooth. Do not allow 
the insert to remain in a static position on any one 
area  of  the  tooth.  Change  the  insert's  path  of 
motion. 
‧ Applying pressure. Use extremely light grasp and 
pressure, especially on exposed cementum. 
‧ If  sensitivity  persists,  decrease  power  setting 
and/or move from the sensitive tooth to another 
and then return.